Determination of copper species in atmospheric waters
Date
2014Author
Subject
Metadata
Show full item recordAbstract
The total copper concentration was determined in rainwater and roof runoff collected in Plymouth, UK between Sept-Oct 2013. The concentration was measured using inductively coupled plasma mass spectrometry and found to be 2.19 ± 0.456 and 3.18 ± 0.506 μg L-1 for rainwater and roof runoff, respectively. These were found to coincide with other published rainwater Cu concentrations. The method itself showed poor accuracy, with a 77.6% recovery for Cu. A method for the determination of Cu(I) by ultraviolet-visible spectrophotometry using bathocuproine was also evaluated. A series of Cu(I)Br calibration standards between concentrations of 0.05 and 0.5 mM were used to assess the absorbance of the [Cu(BCP)2]+ complex at a various pH ranges. Ethylenediamine was used to mask inferences caused by Cu(II). The [Cu(BCP)2]+ complex was observed to be most stable as a pH >8.0, resulting in the highest absorbance throughout the pH ranges. Opened and closed systems were used to evaluate the oxidation of Cu(I) in the presence of air. The open system saw a small decrease in absorbance, inferring the oxidation rate of Cu(I) was not rapid, with only a partial Cu(I) oxidation. The limit of detection for Cu(I) using this method was found to be 1.11 x 10-4 M, concluding that the method was unsuitable for quantifying Cu(I) in atmospheric waters.
Citation
Palmer, J. (2014) 'Determination of copper species in atmospheric waters', The Plymouth Student Scientist, 7(2), p. 151-184.
Publisher
Journal
Volume
Issue
Recommended, similar items
The following license files are associated with this item: